The FIA680 and FIA880 are dedicated fluorescence immunoassay analyzers that read Migibio FICT test cartridges and convert fluorescence signal into quantitative concentration results.

How Fluorescence Detection Works

FICT (Fluorescence Immunochromatographic Technology) is instrument-read, not visually read:

LED excitation → fluorophore emission → photodetector → fluorescence intensity
→ standard curve → concentration
  1. Excitation — an LED excites the fluorophore bound to the detection antibody at the test line.
  2. Emission — the fluorophore emits light at a longer wavelength (Stokes shift).
  3. Detection — a photodiode (or PMT) measures the emitted light intensity.
  4. Quantification — intensity maps to analyte concentration via the pre-loaded standard curve.

This is fundamentally different from colloidal-gold (visual) strips: fluorescence intensity is measured quantitatively, giving concentration values rather than a "line present / absent" judgment.

Analyzer Comparison

Parameter FIA680 FIA880
Channels Single Six
Throughput 1 test/run Up to 6 tests/run
Sample types Serum, plasma, whole blood Serum, plasma, whole blood
Excitation LED LED
Detection Photodiode Photodiode / PMT
Calibration QR code standard curve Cloud standard-curve sync
Best for Low-volume clinics High-volume clinics

Operation Overview

  1. Apply sample to the test cartridge.
  2. Insert cartridge into the analyzer.
  3. Read result — concentration + qualitative flag in 3–15 min.

Calibration & Maintenance

Item Frequency
QC cartridge verification Daily (or per IFU)
Standard-curve update Via QR code / cloud sync
Optical path cleaning Weekly
Full calibration Per service schedule

FAQ

Do I need to build a standard curve manually? No — curves are pre-loaded via QR code or cloud sync.

Can I run different assays simultaneously? On FIA880 yes (six channels, mixed assays); FIA680 is single-channel.

What samples are accepted? Serum, plasma, and whole blood (assay-dependent).

Why is fluorescence better than visual (colloidal gold) reading? Fluorescence gives quantitative concentrations (pg/ml sensitivity, CV% < 10%), while visual strips are qualitative and operator-dependent. See Quantitative vs Qualitative.

For assay performance data, see the Data Hub.

Authored by: Migibio Clinical & Scientific Affairs, Guangzhou Magic Biotech Co., Ltd.

Reviewed by: Migibio R&D Quality Committee

Last updated: 2026-08-14

Disclosure: Migibio (Guangzhou Magic Biotech Co., Ltd.) is the manufacturer of the FIA680/FIA880 analyzers and FICT reagents referenced in this content. See our Editorial & Review Policy.

Contact: Martin.Wong  ·  Phone (WhatsApp): +86 13323237275  ·  Email: [email protected]
Back to top