Calibration verification confirms that the analyzer — and its loaded standard curve — produces correct results. This page covers what to verify, how often, and why.

Calibration vs Verification

Term Meaning
Calibration Establishing the curve (done at manufacturing; loaded via QR/cloud)
Verification Confirming the loaded curve reads correctly on your instrument

The operator does not calibrate — they verify that the pre-loaded curve is reading correctly.

Verification Methods

Method What it confirms
QC cartridge (known value) Analyzer reads the expected control concentration
Control material Curve accuracy at a defined concentration
Proficiency sample Agreement with external reference

Verification Frequency

Trigger Action
Daily QC cartridge verification
New assay lot Verify the new lot's curve
Instrument service/repair Full verification
Suspicion of drift Immediate verification

Acceptance Criteria

Parameter Acceptance
QC result Within the control's stated range
Curve integrity No error on load
Inter-operator agreement Consistent results across operators

FAQ

Do I ever build a standard curve myself? No — curves are pre-loaded (QR code or cloud). Your role is verification, not construction.

What if my QC value drifts within range? Monitor the trend; consistent drift toward the limit warrants proactive service.

For the QC program, see Internal QC Program.

Authored by: Migibio Clinical & Scientific Affairs, Guangzhou Magic Biotech Co., Ltd.

Reviewed by: Migibio R&D Quality Committee

Last updated: 2026-08-13

Disclosure: Migibio (Guangzhou Magic Biotech Co., Ltd.) is the manufacturer of the FIA680/FIA880 analyzers and FICT reagents referenced in this content. See our Editorial & Review Policy.

Contact: Martin.Wong  ·  Phone (WhatsApp): +86 13323237275  ·  Email: [email protected]
Back to top